1
Materials and Methods
1
1
1
Instruments and Reagents
Agilent Technologies 7890A - 5975C Gas Chromatography-Mass Spectrometry (United States, Agilent); Teledyne Tekmar Velocity XPT Purge Trap (United States, Tekmar), 5 ml Purge Tube (United States, Tekmar); Electronic Balance (Switzerland, METTLER, accurate to 0.1 mg); Milli-Q Ultra-pure Water System (United States, Millipore); Vials, 40 ml, baked at 180°C for at least 2 hours before use; Micro-syringe, 10 and 100 µl; Chromatographic-grade methanol; Milli-Q ultra-pure water; Standard material: 502/524 VOC Mix, containing 54 VOCs, concentration 2000 µg/ml (Supelco, United States); Internal standard: Fluorobenzene (Dr. Ehrenstorfer, Germany, 99.5%); Carrier gas: High-purity helium; Purge gas: High-purity nitrogen.
1
1
2
Measuring Conditions
1.2.1 Chromatographic Conditions: Column DB-WAX (20 m × 0.18 mm × 1.0 µm), pressure: 20 psi; column flow: 1.10 ml/min; column temperature: 40°C for 3 min, 10°C/min to 100°C, 0 min hold, 25°C/min to 225°C, 3 min hold; inlet temperature: 150°C; split ratio: 50:1.
1.2.2 Mass Spectrometry Interface Temperature: 280°C; Ion Source Temperature: 230°C; Quadrupole Temperature: 150°C; EM Voltage: 1400 V; Solvent Delay: 0 min.
1.2.3 Purge and Trap Conditions: Purge sample volume: 5–10 ml, purge time: 11 min, purge flow: 40 ml/min, desorption temperature: 250°C, desorption flow: 300 ml/min, desorption time: 2 min.
1
1
3
Standard Preparation and Analysis
Accurately weigh fluorobenzene internal standard, prepare a stock solution, dilute with methanol to 20–100 µg/ml. Take 2000 µg/ml VOC mixed standard, dilute to 10–100 µg/ml. Prepare standard series at 0, 0.150, 1.0, 2.0, 5.0, 10.0, 20.0, 50.0, 100.0 µg/L. Add 5–10 µl of internal standard solution to each 40 ml vial filled with ultrapure water, analyze using purge and trap-GC-MS. Standards should be freshly prepared daily.
1
1
4
Sample Collection and Analysis
Bake 40 ml sample bottles, fill with water without air gaps, add 25 mg ascorbic acid, seal at 4°C, and analyze as soon as possible. Before analysis, add 5–10 µl of internal standard solution to each sample, purge, adsorb, and analyze using GC-MS.
2
Results and Discussion
2.1 Experimental Parameter Determination: Using the internal standard method, plot the peak area ratio of the standard quantification ion to the internal standard against concentration to create a calibration curve. As shown in Table 1, 28 VOCs showed good linearity (R² > 0.996) in the range of 0–50 µg/L. Six replicate samples at 20–100 µg/L had RSD between 2.12% and 13.0%. Detection limits were between 0.001 and 0.01 µg/L.
2.1.2 Optimization of Experimental Conditions: Compared to other columns like HP-VOC or DB-624, the DB-WAX column (20 m × 0.18 mm × 1.0 µm) provided better separation of 28 VOCs in 14 minutes. Although xylene and bromoform had similar retention times, their quantification ions differ, so no interference occurred. The mixed standard total ion chromatogram is shown in Figure 1.
2.1.2.2 Purge and Trap Conditions: The VOCARB®3000 trap was chosen for its high capture and release efficiency. Optimal conditions included a purge time of 11 min, purge flow of 40 ml/min, desorption temperature of 250°C, desorption flow of 300 ml/min, and desorption time of 2 min. These settings resulted in a recovery rate of 85.5%–108.6% for tap water samples.
2.1.3 Determination of Actual Samples: A city’s tap water was tested, and VOCs like chloroform, bromoform, and styrene were detected at concentrations between 0.003 and 0.040 mg/L. Spiked samples showed recoveries of 85.5%–108.6%, confirming the method’s reliability.
3
Conclusion
This study established a rapid method for determining 28 VOCs in tap water using purge and trap-GC-MS. The method is simple, fast, and effective. A preliminary investigation into VOC levels in a city's tap water revealed increasing contamination due to industrial waste, urban sewage, and agricultural runoff. These contaminants not only persist during water treatment but also react with disinfectants to form harmful by-products. To ensure safe drinking water, stricter regulations, improved disinfection technologies, and enhanced environmental controls are essential to reduce health risks.
The plate-grid is a kind of Printer Accessories.
The cartridge is irradiated by laser beam to adsorb toner, and then the toner is hot pressed by fixing roller for printing. In this process, there will be part of toner residual, which can not be "granules returned to the warehouse"; Automatic cleaning function is not adsorbed new toner particles and directly print, will remain toner away, fully ensure the next printing effect. And the plate-grid plays an important role. When high voltage generator to a high voltage electrode, wire electrode with reseau formed between a strong electric field, and release the corona, wire electrode and the photosensitive drum ionizes the air between the air ions migrate to the drum surface, make the photoconductor (drum) surface is full of charge, so can spare toner "adsorption to warehouse", so as to save toner, The purpose of reducing environmental pollution.
Printer Plate-Grid is widely used in high-end Printer, photocopier, laser printer cartridge. Our charging network plate grid is made by SUS304 stainless steel. We use fine metal etching process, and we can guarantee that the etched plate grid has dense charging network, small diameter, etching straight lines, no broken lines and no notch. We custom Plate-Grid with drawings provided by customers.
Custom Plate Grid,Etched Plate Grid,Printer Plate-Grid,Charging Network Plate Grid
SHAOXING HUALI ELECTRONICS CO., LTD. , https://www.cnsxhuali.com